Type N thermocouple table: mV vs °C (−270 to 1300 °C)
Type N thermocouple voltage in mV from −270 to 1300 °C in 10 °C steps, from the NIST (ITS-90) reference functions with the reference junction at 0 °C.
A type N thermocouple (nickel-chromium-silicon (+) and nickel-silicon (−)) produces a voltage that depends on the temperature difference between the measuring junction and the reference junction. The table follows the NIST (ITS-90) reference functions, with the reference junction at 0 °C.
| °C | mV |
|---|---|
| −270 | −4.345 |
| −260 | −4.336 |
| −250 | −4.313 |
| −240 | −4.277 |
| −230 | −4.226 |
| −220 | −4.162 |
| −210 | −4.083 |
| −200 | −3.990 |
| −190 | −3.884 |
| −180 | −3.766 |
| −170 | −3.634 |
| −160 | −3.491 |
| −150 | −3.336 |
| −140 | −3.171 |
| −130 | −2.994 |
| −120 | −2.808 |
| −110 | −2.612 |
| −100 | −2.407 |
| −90 | −2.193 |
| −80 | −1.972 |
| −70 | −1.744 |
| −60 | −1.509 |
| −50 | −1.269 |
| −40 | −1.023 |
| −30 | −0.772 |
| −20 | −0.518 |
| −10 | −0.260 |
| 0 | 0.000 |
| 10 | 0.261 |
| 20 | 0.525 |
| 30 | 0.793 |
| 40 | 1.065 |
| 50 | 1.340 |
| 60 | 1.619 |
| 70 | 1.902 |
| 80 | 2.189 |
| 90 | 2.480 |
| 100 | 2.774 |
| 110 | 3.072 |
| 120 | 3.374 |
| 130 | 3.680 |
| 140 | 3.989 |
| °C | mV |
|---|---|
| 150 | 4.302 |
| 160 | 4.618 |
| 170 | 4.937 |
| 180 | 5.259 |
| 190 | 5.585 |
| 200 | 5.913 |
| 210 | 6.245 |
| 220 | 6.579 |
| 230 | 6.916 |
| 240 | 7.255 |
| 250 | 7.597 |
| 260 | 7.941 |
| 270 | 8.288 |
| 280 | 8.637 |
| 290 | 8.988 |
| 300 | 9.341 |
| 310 | 9.696 |
| 320 | 10.054 |
| 330 | 10.413 |
| 340 | 10.774 |
| 350 | 11.136 |
| 360 | 11.501 |
| 370 | 11.867 |
| 380 | 12.234 |
| 390 | 12.603 |
| 400 | 12.974 |
| 410 | 13.346 |
| 420 | 13.719 |
| 430 | 14.094 |
| 440 | 14.469 |
| 450 | 14.846 |
| 460 | 15.225 |
| 470 | 15.604 |
| 480 | 15.984 |
| 490 | 16.366 |
| 500 | 16.748 |
| 510 | 17.131 |
| 520 | 17.515 |
| 530 | 17.900 |
| 540 | 18.286 |
| 550 | 18.672 |
| 560 | 19.059 |
| °C | mV |
|---|---|
| 570 | 19.447 |
| 580 | 19.835 |
| 590 | 20.224 |
| 600 | 20.613 |
| 610 | 21.003 |
| 620 | 21.393 |
| 630 | 21.784 |
| 640 | 22.175 |
| 650 | 22.566 |
| 660 | 22.958 |
| 670 | 23.350 |
| 680 | 23.742 |
| 690 | 24.134 |
| 700 | 24.527 |
| 710 | 24.919 |
| 720 | 25.312 |
| 730 | 25.705 |
| 740 | 26.098 |
| 750 | 26.491 |
| 760 | 26.883 |
| 770 | 27.276 |
| 780 | 27.669 |
| 790 | 28.062 |
| 800 | 28.455 |
| 810 | 28.847 |
| 820 | 29.239 |
| 830 | 29.632 |
| 840 | 30.024 |
| 850 | 30.416 |
| 860 | 30.807 |
| 870 | 31.199 |
| 880 | 31.590 |
| 890 | 31.981 |
| 900 | 32.371 |
| 910 | 32.761 |
| 920 | 33.151 |
| 930 | 33.541 |
| 940 | 33.930 |
| 950 | 34.319 |
| 960 | 34.707 |
| 970 | 35.095 |
| 980 | 35.482 |
| °C | mV |
|---|---|
| 990 | 35.869 |
| 1000 | 36.256 |
| 1010 | 36.641 |
| 1020 | 37.027 |
| 1030 | 37.411 |
| 1040 | 37.795 |
| 1050 | 38.179 |
| 1060 | 38.562 |
| 1070 | 38.944 |
| 1080 | 39.326 |
| 1090 | 39.706 |
| 1100 | 40.087 |
| 1110 | 40.466 |
| 1120 | 40.845 |
| 1130 | 41.223 |
| 1140 | 41.600 |
| 1150 | 41.976 |
| 1160 | 42.352 |
| 1170 | 42.727 |
| 1180 | 43.101 |
| 1190 | 43.474 |
| 1200 | 43.846 |
| 1210 | 44.218 |
| 1220 | 44.588 |
| 1230 | 44.958 |
| 1240 | 45.326 |
| 1250 | 45.694 |
| 1260 | 46.060 |
| 1270 | 46.425 |
| 1280 | 46.789 |
| 1290 | 47.152 |
| 1300 | 47.513 |
What about the cold junction?
In the field the reference junction sits at the transmitter or instrument terminals, almost never at 0 °C. So the voltage measured at the terminals is not the table value: the instrument compensates the cold junction. To check a mV reading, take the terminal temperature into account.
Calculated reference values, not measurements. The thermocouple tolerance depends on its class (IEC 60584-1) and is not in the table.
Values calculated from the reference functions; they do not replace measurement or a calibration certificate.