Type K thermocouple table: mV vs °C (−270 to 1372 °C)
Type K thermocouple voltage in mV from −270 to 1372 °C in 10 °C steps, from the NIST (ITS-90) reference functions with the reference junction at 0 °C.
A type K thermocouple (nickel-chromium (+) and nickel-aluminium (−)) produces a voltage that depends on the temperature difference between the measuring junction and the reference junction. The table follows the NIST (ITS-90) reference functions, with the reference junction at 0 °C.
| °C | mV |
|---|---|
| −270 | −6.458 |
| −260 | −6.441 |
| −250 | −6.404 |
| −240 | −6.344 |
| −230 | −6.262 |
| −220 | −6.158 |
| −210 | −6.035 |
| −200 | −5.891 |
| −190 | −5.730 |
| −180 | −5.550 |
| −170 | −5.354 |
| −160 | −5.141 |
| −150 | −4.913 |
| −140 | −4.669 |
| −130 | −4.411 |
| −120 | −4.138 |
| −110 | −3.852 |
| −100 | −3.554 |
| −90 | −3.243 |
| −80 | −2.920 |
| −70 | −2.587 |
| −60 | −2.243 |
| −50 | −1.889 |
| −40 | −1.527 |
| −30 | −1.156 |
| −20 | −0.778 |
| −10 | −0.392 |
| 0 | 0.000 |
| 10 | 0.397 |
| 20 | 0.798 |
| 30 | 1.203 |
| 40 | 1.612 |
| 50 | 2.023 |
| 60 | 2.436 |
| 70 | 2.851 |
| 80 | 3.267 |
| 90 | 3.682 |
| 100 | 4.096 |
| 110 | 4.509 |
| 120 | 4.920 |
| 130 | 5.328 |
| 140 | 5.735 |
| °C | mV |
|---|---|
| 150 | 6.138 |
| 160 | 6.540 |
| 170 | 6.941 |
| 180 | 7.340 |
| 190 | 7.739 |
| 200 | 8.138 |
| 210 | 8.539 |
| 220 | 8.940 |
| 230 | 9.343 |
| 240 | 9.747 |
| 250 | 10.153 |
| 260 | 10.561 |
| 270 | 10.971 |
| 280 | 11.382 |
| 290 | 11.795 |
| 300 | 12.209 |
| 310 | 12.624 |
| 320 | 13.040 |
| 330 | 13.457 |
| 340 | 13.874 |
| 350 | 14.293 |
| 360 | 14.713 |
| 370 | 15.133 |
| 380 | 15.554 |
| 390 | 15.975 |
| 400 | 16.397 |
| 410 | 16.820 |
| 420 | 17.243 |
| 430 | 17.667 |
| 440 | 18.091 |
| 450 | 18.516 |
| 460 | 18.941 |
| 470 | 19.366 |
| 480 | 19.792 |
| 490 | 20.218 |
| 500 | 20.644 |
| 510 | 21.071 |
| 520 | 21.497 |
| 530 | 21.924 |
| 540 | 22.350 |
| 550 | 22.776 |
| 560 | 23.203 |
| °C | mV |
|---|---|
| 570 | 23.629 |
| 580 | 24.055 |
| 590 | 24.480 |
| 600 | 24.905 |
| 610 | 25.330 |
| 620 | 25.755 |
| 630 | 26.179 |
| 640 | 26.602 |
| 650 | 27.025 |
| 660 | 27.447 |
| 670 | 27.869 |
| 680 | 28.289 |
| 690 | 28.710 |
| 700 | 29.129 |
| 710 | 29.548 |
| 720 | 29.965 |
| 730 | 30.382 |
| 740 | 30.798 |
| 750 | 31.213 |
| 760 | 31.628 |
| 770 | 32.041 |
| 780 | 32.453 |
| 790 | 32.865 |
| 800 | 33.275 |
| 810 | 33.685 |
| 820 | 34.093 |
| 830 | 34.501 |
| 840 | 34.908 |
| 850 | 35.313 |
| 860 | 35.718 |
| 870 | 36.121 |
| 880 | 36.524 |
| 890 | 36.925 |
| 900 | 37.326 |
| 910 | 37.725 |
| 920 | 38.124 |
| 930 | 38.522 |
| 940 | 38.918 |
| 950 | 39.314 |
| 960 | 39.708 |
| 970 | 40.101 |
| 980 | 40.494 |
| °C | mV |
|---|---|
| 990 | 40.885 |
| 1000 | 41.276 |
| 1010 | 41.665 |
| 1020 | 42.053 |
| 1030 | 42.440 |
| 1040 | 42.826 |
| 1050 | 43.211 |
| 1060 | 43.595 |
| 1070 | 43.978 |
| 1080 | 44.359 |
| 1090 | 44.740 |
| 1100 | 45.119 |
| 1110 | 45.497 |
| 1120 | 45.873 |
| 1130 | 46.249 |
| 1140 | 46.623 |
| 1150 | 46.995 |
| 1160 | 47.367 |
| 1170 | 47.737 |
| 1180 | 48.105 |
| 1190 | 48.473 |
| 1200 | 48.838 |
| 1210 | 49.202 |
| 1220 | 49.565 |
| 1230 | 49.926 |
| 1240 | 50.286 |
| 1250 | 50.644 |
| 1260 | 51.000 |
| 1270 | 51.355 |
| 1280 | 51.708 |
| 1290 | 52.060 |
| 1300 | 52.410 |
| 1310 | 52.759 |
| 1320 | 53.106 |
| 1330 | 53.451 |
| 1340 | 53.795 |
| 1350 | 54.138 |
| 1360 | 54.479 |
| 1370 | 54.819 |
| 1372 | 54.886 |
What about the cold junction?
In the field the reference junction sits at the transmitter or instrument terminals, almost never at 0 °C. So the voltage measured at the terminals is not the table value: the instrument compensates the cold junction. To check a mV reading, take the terminal temperature into account.
Calculated reference values, not measurements. The thermocouple tolerance depends on its class (IEC 60584-1) and is not in the table.
Values calculated from the reference functions; they do not replace measurement or a calibration certificate.