Scale calibration sheet: load, repeatability and eccentricity in PDF

Prepare the data collection of the three usual scale tests: indication by load, repeatability and eccentricity. Set capacity, loads, cycles or positions, record the standard weight mass and the indication, and export the sheet as an unbranded A4 PDF.

Subscription tool

Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access or subscribe.

Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access.

Want to check the quality first? Free, no sign-up: 4–20 mA calculator · residential cable sizing (whole house)

Scale collection sheet filled with the example: loads from 0 to 100 kg, readings and differenceEnlargeScale collection sheet filled with the example: loads from 0 to 100 kg, readings and difference
The field sheet filled with the example, as it comes out in the PDF. Real crop from the tool, with the example on this page.

Where it helps at work

  • Check shipping, dosing and laboratory scales in routine maintenance.
  • Record the eccentricity test with the platform positions of your procedure.
  • Repeat the same load in cycles to assess repeatability.
  • Compare the scale before and after adjustment with the same setup.

What you enter

  • Test
  • Capacity and unit
  • Loads, cycles or positions
  • Identification

What you get

  • Field sheet
  • Filled collection
  • Means and difference
  • Draft and backup

Worked example: platform scale with standard weights

A 100 kg platform scale with a 0.02 kg division goes through the load indication test with standard weights. The sheet was prepared with 5 loads, from 0 to 100 kg, and 2 readings per load. P records the conventional mass of the weights applied (from the certificate) and I the scale indication.

Data
TestIndication by load
Capacity considered100 kg
Nominal loads0 · 25 · 50 · 75 · 100 kg
Readings per load2
StandardClass M1 standard weights, 4 × 20 kg and 4 × 5 kg
Result
Load 0 kgP 0.000 · I 0.000 · I − P 0.000 kg
Load 25 kgP 25.001 · I 25.020 · I − P +0.019 kg
Load 50 kgP 50.002 · I 50.030 · I − P +0.028 kg
Load 75 kgP 75.003 · I 75.050 · I − P +0.047 kg
Load 100 kgP 100.004 · I 100.070 · I − P +0.066 kg

How to read it

The scale reads above the applied mass, and the difference grows with load: +0.028 kg at 50 kg and +0.066 kg at capacity, equivalent to 3.3 divisions. The pattern suggests a span calibration error of the scale. The sheet does not apply the maximum permissible error: compare with the scale class criterion and your procedure. For repeatability and eccentricity, the tool prepares their own sheets, with cycles of the same load or positions on the platform.

Limits and cautions

  • The sheet records data: it does not apply the maximum permissible error, estimate uncertainty or decide conformity or legal metrology approval.
  • The conventional mass of the weights and its corrections come from the certificate and are typed by you.
  • Discrimination, creep and tare tests have no sheet of their own; use the notes when the procedure asks for them.
  • The draft stays in this browser; use the backup file to keep collections.

Technical basis

Load indication, repeatability and eccentricity tests, in line with OIML R 76 and EURAMET cg-18.

  • OIML R 76-1 — non-automatic weighing instruments
  • EURAMET cg-18 — calibration of non-automatic weighing instruments
  • OIML R 111-1 — weights of classes E₁ to M₃
How we verify the calculations

In the full version, you can

  • Prepare the load, repeatability and eccentricity tests.
  • Record the weight mass and the indication and see the difference per row.
  • Export the blank or filled sheet as an unbranded A4 PDF.
  • Keep the draft, duplicate the setup and save a backup file.

ALOGY Tools subscription

One plan unlocks every subscription tool, with PDF reports and work kept in your account, to open on your phone and computer.

  • Every subscription tool, in Portuguese, English and Spanish
  • Pix or card on Asaas’s secure page, no boleto
  • 7 days to withdraw with a full refund; after that, cancel renewal whenever you want
Log in to my account

Frequently asked questions

What is the difference between d and e?

d is the indication division, the smallest step the scale shows. e is the verification interval, used in the maximum error rules of regulated scales. They may or may not be equal; the sheet records both.

Why is P not the nominal value of the weight?

Because each standard weight has a conventional mass and an error in its certificate. The nominal load only organizes the test; P is the mass actually applied.

How do I record eccentricity?

Choose the eccentricity test, enter the load and the positions of your procedure (centre and corners, for example) and record the indication at each position.