Calibration · pressureSubscription
Pressure calibration sheet: rising, falling and field sheet in PDF
Prepare the data collection of a gauge, transmitter or pressure switch calibration: range, pressure type (gauge, absolute or differential), points, repetitions and direction. Record P and I on the device or print the field sheet, and export it as an unbranded A4 PDF.
Subscription tool
Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access or subscribe.
Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access.
Want to check the quality first? Free, no sign-up: 4–20 mA calculator · residential cable sizing (whole house)
EnlargeWhere it helps at work
- Calibrate gauges and transmitters by comparison with a reference on a hand pump or deadweight tester.
- Record rising and falling to see the hysteresis.
- Take the printed sheet to the shutdown and enter the data later.
- Record before and after the adjustment in the same sheet and see the error of each point side by side.
- Start the next collection of the same TAG from the last saved sheet.
What you enter
What you get
Worked example: gauge with hysteresis
A 0 to 10 bar Bourdon gauge is compared with a digital reference gauge on a calibration pump. The sheet was prepared with 5 points, applied with rising and then falling pressure, with one reading on each pass.
| Working range | 0 … 10 bar |
|---|---|
| Pressure type | Gauge |
| Nominal points | 0 · 2.5 · 5 · 7.5 · 10 bar |
| Direction | Rising and falling |
| Readings per point | 1 |
| Comparison method | Comparison with a digital reference gauge on a calibration pump |
| 0 bar, rising | P 0.00 · I 0.00 · I − P 0.00 bar |
|---|---|
| 2.5 bar, rising | P 2.50 · I 2.54 · I − P +0.04 bar |
| 5 bar, rising | P 5.01 · I 5.07 · I − P +0.06 bar |
| 7.5 bar, rising | P 7.50 · I 7.55 · I − P +0.05 bar |
| 10 bar, rising | P 10.00 · I 10.04 · I − P +0.04 bar |
| 10 bar, falling | P 10.00 · I 10.04 · I − P +0.04 bar |
| 7.5 bar, falling | P 7.50 · I 7.61 · I − P +0.11 bar |
| 5 bar, falling | P 5.00 · I 5.13 · I − P +0.13 bar |
| 2.5 bar, falling | P 2.50 · I 2.61 · I − P +0.11 bar |
| 0 bar, falling | P 0.00 · I 0.06 · I − P +0.06 bar |
How to read it
At 5 bar, rising, the gauge reads +0.06 bar against the reference; falling, at the same point, +0.13 bar. That difference between directions is hysteresis, typical of friction and backlash in the Bourdon mechanism, and it only shows because the sheet records rising and falling. At 0 bar, falling, the pointer did not return to zero (+0.06 bar). Compare the values with the class criterion and your procedure before deciding on adjustment or replacement.
Limits and cautions
- The sheet records data: it does not estimate uncertainty, calculate hysteresis or decide conformity, and does not replace a certificate.
- The liquid head between standard and instrument and other corrections must be handled by you before recording P.
- Simulating the signal tests the electronics; it does not calibrate the sensing element.
- The draft stays in this browser; use the backup file to keep collections.
Technical basis
Data collection sheet for calibrating pressure instruments by comparison, rising and falling, in line with EURAMET cg-17 and ISO/IEC 17025.
- ISO/IEC 17025 — technical records of calibration
- EURAMET cg-17 — calibration of electromechanical pressure instruments
- EN 837-1 — Bourdon tube pressure gauges
In the full version, you can
- Set range, pressure type, points, repetitions and direction.
- Record readings and see means and difference per point and direction.
- Export the blank or filled sheet as an unbranded A4 PDF.
- Keep the draft, duplicate the setup and save a backup file.
ALOGY Tools subscription
One plan unlocks every subscription tool, with PDF reports and work kept in your account, to open on your phone and computer.
- Every subscription tool, in Portuguese, English and Spanish
- Pix or card on Asaas’s secure page, no boleto
- 7 days to withdraw with a full refund; after that, cancel renewal whenever you want
Frequently asked questions
Why record rising and falling?
Because many instruments read differently as pressure rises or falls, due to friction or backlash. Without both directions that hysteresis stays hidden.
Does the height between standard and instrument matter?
Yes, with liquid in the line. Each metre of water column is worth about 0.098 bar. Record the height reference in the identification and correct P when the procedure asks.
Does it work for pressure switches?
As a record of the comparison. For pressure switches, write the trip pressure in P and the set value in I, and use the point notes for the contact state.