Flow calibration sheet: reference method, points and sheet in PDF
Prepare the data collection of a flowmeter calibration by comparison with a reference meter, volumetric or gravimetric collection. Set range, unit, points and repetitions, record P and I and, when the method asks, the volume or mass and time of each collection. Export as an unbranded A4 PDF.
Subscription tool
Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access or subscribe.
Read the explanation and the example freely. To use the tool with your own data, log in to an account with active access.
Want to check the quality first? Free, no sign-up: 4–20 mA calculator · residential cable sizing (whole house)
EnlargeWhere it helps at work
- Compare a process meter with a master meter installed in series.
- Record a tank or scale collection with the volume, mass and time of each run.
- Document the check of internal billing or dosing meters.
- Repeat the collection after adjusting the meter factor.
What you enter
What you get
Worked example: electromagnetic meter against a master meter
A 0 to 100 m³/h electromagnetic flowmeter is compared in series with a master meter. The sheet was prepared with manual points at 25; 50; 75; 100 m³/h and 2 readings per point, recording the master flow (P) and the meter indication (I).
| Range | 0 … 100 m³/h |
|---|---|
| Reference method | Series comparison with a master meter on the water line |
| Points (manual) | 25; 50; 75; 100 |
| Readings per point | 2 |
| Fluid | Water at 22 °C |
| 25 m³/h | P 25.10 · I 25.29 · I − P +0.19 m³/h |
|---|---|
| 50 m³/h | P 50.01 · I 50.19 · I − P +0.18 m³/h |
| 75 m³/h | P 75.06 · I 75.26 · I − P +0.19 m³/h |
| 100 m³/h | P 99.99 · I 100.24 · I − P +0.25 m³/h |
How to read it
The difference was almost constant, around +0.2 m³/h over the whole range, which looks like a zero offset. As a % of reading (calculated here, outside the sheet), it weighs more at low flow: 0.76 % at 25 m³/h against 0.25 % at 100 m³/h. Since many meters state accuracy as % of reading, the low point is the one that decides. The sheet does not pass or fail: compare with the procedure and manufacturer criterion.
Limits and cautions
- The sheet records data: it does not calculate the reference flow from volume, mass and time, does not estimate uncertainty and does not decide conformity.
- Density, reference conditions and standard corrections are handled by your procedure before recording P.
- Simulating the electrical signal tests the electronics, but does not calibrate the process flow measurement.
- The draft stays in this browser; use the backup file to keep collections.
Technical basis
Data collection sheet for flowmeter calibration by comparison, volumetric or gravimetric collection (ISO 4185, ISO 8316).
- ISO 4185 — measurement of liquid flow: weighing method
- ISO 8316 — measurement of liquid flow: volumetric collection
- ISO/IEC 17025 — technical records of calibration
In the full version, you can
- Set range, unit, reference method, points and repetitions.
- Record P, I and the auxiliary data and see the difference per point.
- Export the blank or filled sheet as an unbranded A4 PDF.
- Keep the draft, duplicate the setup and save a backup file.
ALOGY Tools subscription
One plan unlocks every subscription tool, with PDF reports and work kept in your account, to open on your phone and computer.
- Every subscription tool, in Portuguese, English and Spanish
- Pix or card on Asaas’s secure page, no boleto
- 7 days to withdraw with a full refund; after that, cancel renewal whenever you want
Frequently asked questions
Does the sheet calculate flow from volume and time?
No. Volume or mass and time are recorded, but P is the flow your procedure determined, with the density and condition corrections it requires.
Why use manual points?
Because many meters are checked at the flows they actually work at. The manual list accepts up to 50 points within the range.
Can I work in mass, such as kg/h?
Yes. The sheet unit can be mass or volume; P and I must be in the same unit.